THERMOCOUPLES
INTRODUCTION
The basic theory of thermocouples dates
back to 1821 when T.J. Seebeck discovered that a current is
induced into a closed circuit of two dissimilar metals by
heating one of the two junctions. And, as long as the temperature
differences exists between the two junctions, current will
continue flowing through the circuit.
While the theory is nearly 150 years
old, incorrect application of thermocouples still affects
today's sophisticated industrial processes. In any temperature
control system, the heart of that system is the temperature
sensing device -- in this case, thermocouple. Without proper
application or understanding of basic thermocouple circuits,
even the most complicated system cannot function.
In his discovery, Seebeck also concluded
that any two metals can be used. However, the magnitude and
direction of the generated current are functions of the magnitude
of the temperature difference between the junctions and the
thermal properties of the metals used in the circuit. Therefore,
not every combination of metals is acceptable for thermocouple
usage.
THERMOELECTRIC CHARACTERISTICS
A thermocouple should have thermoelectric
characteristics such that the electromotive force (emf) produced
per degree of temperature change is sufficient to be detected
by standard measuring instruments. The device must also be
capable of withstanding temperature extremes for prolonged
periods, rapid temperature changes, and corrosive atmospheres
while exhibiting reproducibility and a high degree of accuracy.
The Instrument Society of America (ISA)
has established a type of code and limits-of-error specifications
for thermocouple wire, shown below:
Table 1: Thermocouple Wire Specifications
|
|
|
ISA LIMITS OF ERROR |
| DESCRIPTION |
ISA TYPE |
TEMPERATURE RANGE |
STANDARD |
SPECIAL |
| Copper/Constantan |
T |
-300 deg. F to -75 deg. F |
-- |
+/- 1% |
| |
|
-150 deg. F to -75 deg. F |
+/- 2% |
+/- 1% |
| |
|
-75 deg. F to +200 deg. F |
+/- 1.5% |
+/- 3/4 deg F |
| Iron/Constantan |
J |
0 deg. F to +530 deg. F |
+/-4 deg F |
+/- 2 deg F |
| |
|
+530 deg F to +1400 deg F |
+/- 0.75% |
+/- 0.375% |
| Chromel/Constantan |
E |
+32 deg. F to +600 deg. F |
+/- 3 deg F |
-- |
| |
|
+600 deg. F to +1600 deg. F |
+/- 0.5% |
-- |
| Chromel/Alumel |
K |
0 deg. F to +530 deg. F |
+/- 4 deg F |
+/- 2 deg F |
| |
|
+530 deg. F to +2300 deg. F |
+/- 0.75% |
+/- 0.375% |
Platinum/Platinum
(+10% Rhodium) |
S |
0 deg. F to +1000 deg. F |
+/- 5 deg F |
-- |
Platinum/Platinum
(+13% Rhodium) |
R |
+1000 deg. F to +2700 deg. F |
+/- 0.5% |
-- |
Six thermocouples are covered by this
system:
- Copper vs. Constantan (ISA Type T)
(Nickel Alloy)
Superior for use at sub zero temperatures.
Withstands corrosion well and is recommended for temperatures
within the range of -300 deg F to +700 deg F.
- Iron vs. Constantan (ISA Type J)
Above 1000 deg F, the rate of oxidation
of the iron wire increases rapidly and the thermocouple
should be enclosed in a protection tube of suitable material.
Protected Iron vs. Constantan thermocouples are recommended
for temperatures up to 1600 deg F.
- Chromel vs. Alumel (ISA Type K) (Nickel
Chromium Alloy) or (Nickel Manganese, Aluminum, Silicon)
These are trade names of Hoskins Manufacturing
Company. Chromel vs. Alumel thermocouples have excellent
characteristics when supplied with protection tubes up
to 2200 deg F.
- Chromel vs. Constantan (ISA Type E)
Although only in limited use in industrial
applications, this type thermocouple has the highest emf
output of any standardized metallic type. They may be
used in oxidizing, inert or reducing atmospheres to 1600
deg F and at sub-zero temperatures they are not subject
to corrosion. Indications are that the future will see
more consideration given to this combination.
- Platinum vs. Platinum-Rhodium
Platinum vs. 90% Platinum + 10% Rhodium
(ISA Type S) and Platinum vs. 87% Platinum + 13% Rhodium
(ISA Type R) are used for temperatures up to approximately
3100 deg F, depending upon the atmosphere. Both types
should always be provided with a high temperature ceramic
protection tube.
Various other combinations of materials
are used for thermocouples throughout industry but with far
less frequency than the six basic types. Combinations of platinum
and rhodium with various percentages of each have been regularly
available for some years. Iridium vs. Iridium with 40, 50
or 60% Rhodium thermocouples have found acceptance in some
high temperature applications up to 3600 deg F.
Proper thermocouple selection is primarily
determined by the temperature range in which its use is intended.
Other factors, such as atmosphere, abrasion, vibration, and
location will determine the type, size, and configuration
of the complete assembly which includes protection tubes and
mounting facilities.
LOCATION
Proper location of the thermocouple is
probably the most important factor in obtaining accurate temperature
control. Thermocouples should be in a position to have a definite
temperature relationship to the heat source and workload.
A good 'rule of thumb' in locating thermocouples is to place
them between the workload and heat source. The thermocouple
should be located 1/3 the distance from the heat source and
2/3 the distance to the workload.
If a thermocouple is located too close
to the heaters, a long warm-up time will result. The thermocouple
will sense the heat before it reaches the workload, and this
means rapid on/off action of the controller. In effect, the
controller is controlling the heater and not the workload.
In rare cases, voltage will be induced into the thermocouple
circuit at high temperatures when located too near the heaters.
When a thermocouple is located too close
to the workload, there is a substantial delay in sensing the
proper control point and the result is overshooting the temperature.
In most cases, it is better to be too close to the heaters
than the workload as once a temperature point is passed, it
becomes difficult to cool the workload unless a forced cooling
system is used. Two thermocouples connected in parallel could
be used, one located near the heaters and the other near the
workload. Both will balance these two factors and provide
closer control.
Another consideration in location is
when locating a thermocouple in a thermocouple well. If it
is not bottomed correctly, located at the bottom of the well,
the thermocouple will be reading the air temperature around
it and not the temperature of the workload.
COMPENSATION
The compensation method used by all millivoltmeter
manufacturers is to attach a bi-metallic spiral to the top
hairspring of the coil suspension system. This spiral is selected
according to the range of the instrument and will deflect
the indicating pointer correspondingly with changes in ambient
temperature. Once ambient is set mechanically, using a zero
adjust screw, it is not necessary to change the setting during
the operation of the instrument. In solid state instruments,
the compensation is achieved electronically by placing a temperature
sensor, such as a thermistor or RTD, at the cold junction
to monitor its temperature. The signal from this sensor is
used to compensate for variations in cold junction temperature.
The automatic compensation for ambient
temperature is sufficient in most industrial applications.
However, in laboratory experiments or critical control situations,
when maximum accuracy is desired, one of two cold junction
compensation methods are used. One method is to place the
cold junction in an agitated ice bath, shown below. The instrument
will then be set at 32 deg F (0 deg C), which is the temperature
of the ice bath.
In the other method, the cold junction
is situated in a precisely controlled temperature above ambient,
as shown below. In this case, ambient compensation is not
necessary. The mechanical zero adjustment is set at the cold
junction temperature being maintained. The normal temperature
being maintained is 150 to 200 degrees F at the cold junction.
In normal applications, if the cold junction
is located too close to the heat source, conduction and radiation
heating will cause inaccurate readings. Errors will also occur
when using copper wire or the wrong thermocouple lead wire.
When copper wire is used, the cold junction in effect remains
at the thermocouple connector block instead of the instrument.
This will cause the instrument to read low in most cases unless
the cold junction and instrument are known to have the same
ambient temperature.
There is one application where cold junction
compensation is not a factor. When two thermocouples are connected
in series opposing, as shown below, a millivoltage is produced
which is the difference in millivolts between the temperature
at both thermocouples. As the difference in degrees between
the two thermocouples is being measured, cold junction compensation
is not necessary.
Each millivolt measuring instrument is
calibrated for both the type of thermocouple being used and
the length and gauge of the lead wire. The thermocouple lead
wire is in effect in series with the thermocouple wire and
the meter movement. Using wrong thermocouple lead wire can
be avoided by simply following the color-coding used by all
manufacturers (Table 2, below). A solid state controller can
be used with up to 100 ohms of external resistance without
having to be recalibrated.
Table 2: Calibration symbols
and color codes for thermocouple and extension wire
| Type |
ISA Symbol |
Positive (+) |
Polarity-Color Code |
Conductor |
Negative (-)
| |
Overall |
| Thermocouple |
J -- Iron |
+ |
White (Magnetic) |
Constantan |
-
| Red |
Brown |
| Extension |
JX -- Iron |
+ |
White |
Constantan |
-
| Red |
Black |
| Thermocouple |
T -- Copper |
+ |
Blue |
Constantan |
-
| Red |
Brown |
| Extension |
TX -- Copper |
+ |
Blue |
Constantan |
-
| Red |
Blue |
| Thermocouple |
E -- Chromel |
+ |
Tan |
Constantan |
-
| Red |
Brown |
| Extension |
EX -- Chromel |
+ |
Tan |
Constantan |
-
| Red |
Brown |
| Thermocouple |
K -- chromel |
+ |
Yellow |
Alumel (Magnetic) |
-
| Red |
Yellow |
| Extension |
KK -- Chromel |
+ |
Yellow |
Alumel |
-
| Red |
Brown |
| Thermocouple |
S -- PT 10% RH |
+ |
-- |
Platinum |
-
| -- |
-- |
| Thermocouple |
R -- PT 13% RH |
+ |
-- |
Platinum |
-
| -- |
-- |
| Extension |
SX -- Copper |
+ |
Black |
Alloy 11 |
-
| Red |
Green |
When a millivoltmeter is calibrated,
a series resistance (commonly called a calibrating spool)
is used between the moving element coil of the instrument
and the thermocouple tip. The resistance of the wire must
be determined and used in the calibration of the instrument.
If the resistance of the thermocouple wire and extension wire
is higher than the instrument is calibrated for, the temperature
readings will be low and if the resistance is lower, the temperature
readings will be high.
Where the thermocouple and extension
wire are a significant portion of the circuit, then we must
also consider the resistance change of the thermocouple wire
at elevated temperatures. It may be necessary to calibrate
instruments at the operating temperature. As an example: 5
feet of .020 dia. platinum vs. platinum 10% Rhodium thermocouple
wire, would have a resistance of 2.3 ohms. At 2500 deg F,
the resistance would be 2.3 x 3.5 ohms, or 8.5 ohms. A millivoltmeter
with a sensitivity of 10 ohms per volt would have an error
of approximately 4% at 2500 deg F. The effects of temperature
on the thermocouple and thermocouple extension wire are shown
in Table 3.
To illustrate the effects of incorrect
lead length calibration on the millivoltmeters, we have charted
the errors that can result for various ranges and thermocouples
by deviating in resistance from the calibrated lead length.
Table 4 is based upon 10 ohms per millivolt sensitivity instrument.
Instruments with less sensitivity would show greater errors.
A meter with a 5 ohm per millivolt sensitivity would have
errors twice as great.
Table 3: Thermocouple resistance
change with temperature
|
Multiplying
factor for various temps; both wires same gauge |
|
200 deg F |
400 deg F |
800 deg F |
1600 deg F |
2500 deg F |
| Iron-Constantan |
1.02 |
1.05 |
1.11 |
1.22 |
---- |
| Chromel Alumel |
1.05 |
1.14 |
1.30 |
1.62 |
2.01 |
| Chromel-Constantan |
1.13 |
1.33 |
1.7 |
2.5 |
---- |
| Plat. 10% RH - Platinum |
1.13 |
1.34 |
1.83 |
2.67 |
3.50 |
| Plat. 13% RH - Platinum |
1.13 |
1.33 |
1.80 |
2.60 |
3.40 |
Table 4: Deviation in Ohms
from calibrated lead length
 |
1 -- 0-2000 deg F C/A (4.02% at 20 Ohms)
2 -- 0-1200 deg F I/C, 0-1600 deg F C/A (4.97% at 20 Ohms)
3 -- 0-800 deg F I/C, 0-600 deg F C/A (7.65% at 20 Ohms)
4 -- 0-500 deg F CU/C, 0-2200 deg F PLT/PLT + 13% RH
0-2400 deg F PLT/PLT + 10% RH (14.32% at 20 Ohms)
5 -- 0-300 deg F I/C, o-350 deg F CU/C (22.15% at 20 Ohms)
|
THERMOCOUPLE CONNECTION
There are two common errors in connection
thermocouple circuits. One is to connect the extension lead
wire completely reversed. In this case, you would receive
a low reading because the reversal causes the emf generated
at the connection of the thermocouple and extension lead wire
to be subtracted from the emf generated by the thermocouple.
A more obvious error is to completely reverse the thermocouple.
The instrument in this case will read downscale with an increase
in temperature.
Some control instruments feature 'thermocouple
break protection' which means that in the event of an open
or broken thermocouple, a small voltage is applied to the
instrument which will cause it to read full scale and turn
off the external circuit. Thus, in the event the thermocouple
breaks because of a mechanical shock or vibration or if it
is over-exposed to extremely high temperature and deterioration
sets in, an unattended process will not overheat because of
the loss of control.
Another consideration in the thermocouple
use is that the leads wires should never run in the same conduit
with electrical lines. This may induce currents in the thermocouple
wire, resulting in instrument errors and poor control. However,
if this cannot be avoided, or if the induced currents are
being picked up at the thermocouple itself, then one side
of the thermocouple lead wire should be grounded through a
1.0 microfared paper capacitor at one of the thermocouple
terminals in the instrument. In emergencies, a direct
ground will sometimes work as well.
Occasionally, because of atmospheric
conditions, corroding may occur on connections which cause
a loss of the millivolt signal. Or, a poor connection between
the lead wire and thermocouple could cause loss of signal.
The gauge size of the wire used in thermocouples
is again dependent upon the application. Usually, when longer
life is required, for the higher temperature ranges, the larger
size wires are chosen. When sensitivity is the prime concern,
the smaller sizes should be used.
GLOSSARY OF TERMS
CALIBRATE - General: to determine the
indication or output of a measuring device with respect to
that of a standard.
CALIBRATE - Thermocouple: to determine
the emf developed by a thermocouple with respect to temperature
established by a standard.
CALIBRATION POINT - General: a specific
value, established by a standard, at which the indication
or output of a measuring device is determined.
CALIBRATION POINT - Thermocouple: a temperature,
established by a standard, at which the emf developed by a
thermocouple is determined.
CELSIUS - The designation of the degree
on the International Practical Temperature Scale. Also used
for the name of the Scale, as "Celsius temperature scale."
Formerly (prior to 1948) called "centigrade."
CENTIGRADE - The designation of the degree
on the International Temperature Scale prior to 1948. (See
Celsius)
COAXIAL THERMOCOUPLE ELEMENT - A thermocouple
element consisting of a thermoelement in wire form, within
a thermoelement in tube form with the two thermoelements insulated
form each other and from the tube except at the measuring
junction.
CONNECTION HEAD - A housing enclosing
a terminal block for an electrical temperature-sensing device
and usually provided with threaded openings for attachment
to a protecting tube and for attachment of conduit.
ELECTROMOTIVE FORCE - (emf) The electrical
potential difference which produces or tends to produce an
electric current.
EXTENSION WIRE - A pair of wires having
such temperature-emf characteristics relative to the thermocouple
with which the wires are intended to be used that, when properly
connected to the thermocouple, the reference junction is transferred
to the other end of the wires.
FAHRENHEIT - The designation of the degree
and the temperature scale used commonly in public life and
engineering circles in English-speaking countries. Related
to the International Practical Temperature Scale by means
of the equation:
- Degrees Fahrenheit = (Degrees Celsius
x 1.8) + 32
- Degrees Celsius = (Degrees Fahrenheit
- 32) / 1.8
FREEZING POINT - The fixed point
between the solid and liquid phases of a material when
approached from the liquid phase under a pressure of one
standard atmosphere (101325 N/m squared). For a pure material,
this is also the melting point.
ICE POINT - The fixed point between
ice and air-saturated water under a pressure of one standard
atmosphere (101325 N/m squared). This temperature is 0
deg C on the International Practical Temperature Scale.
KELVIN - The designation of the thermodynamic
temperature scale and the degree on this scale. This kelvin
scale was defined by the Tenth General Conference on Weights
and Measures in 1954 by assigning the temperature of 273.16
degrees Kelvin to the triple point of water. Also the
degree on the International Practical Kelvin Temperature
Scale.
MELTING POINT - The fixed point between
the solid and liquid phases of a material when approached
from the solid phase under a pressure of one standard
atmosphere (101325 N/m squared). For a pure material,
this is also the freezing point.
PROTECTING TUBE - A tube designed
to enclose a temperature-sensing device and protect it
from the deleterious effects of the environment. It may
provide for attachment to a connection head, but is not
primarily designed for pressure-tight attachment to a
vessel.
RANGE - The region between the limits
within which a quantity is measured. It is expressed by
stating the lower and upper range-values.
REFERENCE JUNCTION - That junction
of a thermocouple which is at a known temperature.
REFERENCE POINT - (liquid-in-glass
thermometer) A temperature at which a thermometer is checked
for changes in bulb volume.
REFRACTORY METAL THERMOCOUPLE - A
thermocouple whose thermoelements have melting points
above that of 60 percent platinum, 40 percent rhodium,
1935 deg C (3515 deg F)
RESISTANCE, INSULATION - (sheathed
thermocouple wire) The measured resistance between wires
or between wires and sheath multiplied by the length of
the wire expressed in megohms (or ohms) per foot (or meter)
of length. (NOTE: The resistance varies inversely with
the length.)
SEEBECK COEFFICIENT - the rate of
change of thermal emf with temperature at a given temperature.
Normally expressed as emf per unit of temperature. synonymous
with thermoelectric power.
SEEBECK EMF - The net emf set up
in a thermocouple under condition of zero current. It
represents the algebraic sum of the Peltier and Thompson
emf. Synonymous with thermal emf.
SHEATHED THERMOCOUPLE - A thermocouple
having its thermoelements, and sometimes its measuring
junction, embedded in ceramic insulation compacted within
a metal protecting tube.
SHEATHED THERMOCOUPLE MATERIAL -
One or more pairs of thermoelements (without measuring
junction(s)) embedded in ceramic insulation compacted
within a metal protecting tube.
THERMOCOUPLE - Two dissimilar thermoelements
so joined as to produce a thermal emf when the junctions
are at different temperatures.
THERMOCOUPLE ASSEMBLY - An assembly
consisting of a thermocouple element and one or more associated
parts such as terminal block, connection head, and protecting
tube.
THERMOCOUPLE ELEMENT - A pair of
bare or insulated thermoelements joined at one end to
form a measuring junction and intended for use as a thermocouple
or as part of a thermocouple assembly.
THERMOCOUPLE (TYPE E, B, J, K, R,
S, OR T) - A thermocouple having an emf-temperature relationship
corresponding to the appropriate letter-designated table
in ASTM Standard E 230, Temperature Electromotive Force
(EMF) Tables for Thermocouples, within the limits of error
specified in that Standard.
THERMOPILE - A number of thermocouples
connected in series, arranged so that alternate junctions
are at the reference temperature and at the measured temperature,
to increase the output for a given temperature difference
between reference and measuring junctions.
THERMOWELL - A closed end reentrant
tube designed for the insertion of a temperature-sensing
element, and provided with means for pressure-tight attachment
to a vessel.
WORKING STANDARD THERMOCOUPLE - A
thermocouple that has had its temperature-emf relationship
determined by reference to a secondary standard of temperature.
|