Home
Line Card
Product Category
Literature
Trade Show Schedule
Service & Calibration
Company Information
 


  
http://www.branom.com
sales@branom.com
800-767-6051
 

Home > Applications

THERMOCOUPLE CALIBRATION PROCEDURE:

The following information is intended to give the reader a review, in some detail, of the equipment requirements and proper techniques needed to accurately calibrate thermocouples and thermocouple materials.

Branom Instrument calibrates thermocouple and thermocouple wire in accordance with one of the following American Society for Testing and Material (ASTM) Standards: E207-88, standard method of Thermal EMF Test of single thermoelement materials by comparison with a secondary standard of similar EMF temperature properties. E220-86, standard method for calibration of thermocouples by comparison techniques.

In general these standards describe the type of temperature source, measuring equipment, standards, and procedures needed to accurately perform a calibration. Each of these elements will now be looked at more closely.

CONTROLLED TEMPERATURE SOURCE:

The temperature source used in the process of calibrating should as a minimum be stable enough to provide a constant temperature (approx.+/-.2 degrees F) for a short length of time (approx.20 mm.) at any temperature at which the temperature bath or other source is to be used. The temperature source should have a zone of uniform temperature into which the thermocouple measuring junction may be inserted. The length of the temperature source must be adequate to permit a depth of immersion sufficient to assure that the measuring junction temperature is not affected by a temperature gradient along the thermocouple wires.

CONTROLLED TEMPERATURE SOURCES:

  • Fixed Point: When highly accurate measurements must be made, fixed point cells are used. A fixed point cell consists of a metal sample inside a graphite crucible with a graphite thermometer well submerged into the metal sample. When the metal sample is heated to the freezing point, it will produce a very stable and constant temperature. In order to better understand the operation of fixed point cells, the following definitions are useful.
  • Fixed Point: A reproducible temperature of equilibrium between different phases of a material.
  • Freezing Point: The fixed point between the solid and liquid phases of a material.

REFERENCE JUNCTIONS:

A thermocouple's output is based on the difference in temperature between the measuring junction (hot junction) and the reference junction (cold junction). See Figure A.

REFERENCE JUNCTION TEMPERATURE:

A controlled temperature must be provided in which the reference junction is maintained at a constant chosen temperature. The reference junction temperature should be controlled to a better accuracy than that expected from the thermocouple calibration. The most commonly used reference temperature is 32 degrees F., but other temperatures may be used if desired.

ICE BATH:

One of the most common reference junctions is the ice bath. The ice bath is made up of a mixture of melting shaved ice and water. The ice bath is a convenient and inexpensive way to achieve an ice point, it can be reproduced with ease and with exceptional accuracy. Junctions formed between the thermocouple materials and instrument leads can be simply immersed into the slush mixture, or alternatively glass "U" tubes containing a quantity of mercury approximately 3/4" to 1" depth can be placed into the slush mixture. Quick electrical connection can then be made between thermocouple and instrument leads through the mercury. (Figure B).

Note: An improperly used ice bath can result in serious errors. The largest error which is likely to occur arises due to melting of the ice at the bottom of the bath until the reference junctions are below the ice level and surrounded by water alone. This water may be as much as 7 degrees F above the ice point.

AUTOMATIC ICE POINT:

The automatic ice point is an electrical refrigerated device in which an equilibrium between ice and water is constantly maintained. The change of volume of water in freezing is used to control heat transfer. Some commercially available devices provide wells into which the user may insert reference junctions formed from his own calibrated wire. Others are provided with many reference junction pairs brought out to terminals which the user may connect into his system.

ELECTRONIC COMPENSATION:

This method employs a compensation circuit containing a source of current and a combination of fixed resistors and a temperature sensitive resistor (TSR). This device can be designed to produce similar EMF to that of the thermocouple being calibrated. The Electronic Compensator will make EMF compensations to the thermocouple circuit based in the difference in EMF from 32 to ambient temperature.

MEASURING INSTRUMENTS:

The choice of a specific instrument to use for measuring the thermocouple output will depend on the accuracy required of the calibration being performed. In general, an instrument such as the Fluke 702 calibrator or Altek 422 is sufficient for most thermocouple calibrations.

REFERENCE THERMOMETERS:

The reference thermometer to be used for the comparison calibration of a thermocouple will depend upon the temperature range covered, the accuracy desired, the capabilities, or the preference of the calibration laboratory. The following are different examples of reference thermometers.

PLATINUM RESISTANCE THERMOMETERS:

A standard platinum resistance thermometer (SPRT) is the most accurate standard available, however, it is the most expensive standard, and other standards are acceptable alternatives depending upon the temperature range covered, the accuracy desired, the capabilities, or the preference of the calibration laboratory. The following are different examples of reference thermometers.

LIQUID-IN-GLASS THERMOMETERS:

Liquid-in-glass thermometers are available to cover the range from -300 to 950 degrees Fahrenheit. with an accuracy of from .1 to 3 Fahrenheit depending on the type of thermometer and the width of the range covered. They are relatively inexpensive but they are fragile, and if the highest degree of accuracy of which they are capable is to be achieved, an individual thermometer must cover a very narrow temperature range so that the graduation intervals can be as large as possible. A further disadvantage of the liquid-in-glass thermometer is that because of their fine graduations reading errors are a distinct possibility. Taylor Instruments offers Superior Grade Certified Secondary Reference Thermometers individually or in matched Celsius or Fahrenheit sets, which Branom stocks.

TEST ASSEMBLY PLACEMENT IN THE FURNACE:

Depth of immersion is the most important consideration if accurate calibration results are to be obtained. The depth of immersion must be sufficient to eliminate the effects of heat transfer away from the junction. It is impossible to establish a minimum depth of immersion that would be useable under all circumstances since heat transfer characteristics are dependent on the mass of material being put into the temperature source.

WIRING CONNECTION FROM TEST ASSEMBLY TO READOUT INSTRUMENT.

The actual wiring necessary to connect the test assembly, reference junction and readout instrument will depend on the quantity of thermoelements in the test assembly, the type of reference junction used and whether or not a switching device is used, but the basic requirements are the same. Thermocouple extension wire is used to connect the thermoelements to the reference junction. Copper wires are used between the reference junction and readout instrument.

THERMOCOUPLE WIRE, WIRING PROCEDURE:

Ideally, the samples of the thermocouple material to be calibrated and the standard thermocouple element should be cut long enough so that they reach directly from the temperature source to the reference junction without the need for extension wires. If this is not possible extension wires may be used, but they must be securely connected directly to the test assembly conductors. If extension wires must be used, remove any oxide layer that may be on the surface of the test assembly conductors and attach an extension wire of the same material to each conductor by laying the extension wire alongside the conductors and joining them securely by means of an alligator clip.

THERMOCOUPLE CALIBRATION WIRING PROCEDURE:

When calibrating thermocouples, it is faster and more convenient to use a thermocouple switching box. The extension wires from the thermocouples are placed into one side of the reference junction. Multiple pairs of copper leadwire will exit the reference junction and will be connected to the switch box. One pair of copper leadwires will run from the readout instrument to the thermocouple switch box.

For a more in-depth look at thermocouples and thermocouple calibration the reader is encouraged to read ASTM STP 470, manual on the use of thermocouple in temperature measurement.

JUNCTION LOCATION:

One of the primary advantages of calibrating thermocouple materials against a base-metal standard of similar EMF output is that the sample(s) to be calibrated are welded to the base-metal standard forming a common junction thus achieving good isothermal conditions between the test thermoelement and the standard. Furthermore, because the test thermoelement and the standard produce nominally the same EMF vs. platinum (pt-67) the EMF output changes little over a fairly broad temperature range, thereby reducing the need for precise temperature source control. See Figure C.

MEASUREMENT: Set your controlled temperature source to the specified temperature and allow it to adequately stabilize. Immerse the test assembly into the test temperature medium and provide sufficient time for the test assembly to stabilize. Once the test assembly is stable the EMF generated between the test specimen and the reference standard can be recorded. Avoid soaking the test assembly at temperature for a prolonged period of time, as it can cause permanent changes to occur in the thermoelements.

Once the reading is taken, raise the test temperature to the next higher temperature, first removing the test assembly from the temperature source, or advance the test assembly to the next temperature source. Allow the temperature source and the test assembly to stabilize as before, and take a second set of readings at the new temperature. In all cases take the reading in sequence from the lowest to the highest temperature. A base metal reference standard shall be used for one series of temperature changes only.

ASTM E 220 THERMOCOUPLE CALIBRATION:

The Test thermocouple junction should be located so that it is in intimate contact with the junction of the standard. Without making a radiograph of the thermocouple it is impossible to know exactly where the junction is located. A few generalizations can be made which enables junctions to be located quite closely. First, the cap weld on a metal sheathed thermocouple is normally about as thick as one-half the sheathed diameter. Second, a "U" junction is normally about one-half the sheathed diameter. Using these generalizations, a thermocouple .125" diameter, will have a grounded junction approximately .063" below the tip of the cap. The thermocouple standard should be tied to the thermocouple (s) with a fine gauge wire. The junction of the standard should be bent so that it is in contact or at least very close to the point where it has been calculated that the junction is located. See Figure D.

PLATINUM VERSUS PLATINUM RHODIUM THERMOCOUPLE:

Platinum vs. platinum 10% Rhodium standard thermocouples (ANSI Type S) are exceptionally accurate and stable devices. NIST offers calibration uncertainty of.9 F. from 0 to 1112 F, and 1.26 F. from 1112 to 2012 F. When used as a working standard, understandably due to the high cost of these materials, they cannot be discarded after each use. Consequently, care must be taken to avoid contamination, work hardening and other sources of de-calibration.

BASE METAL THERMOCOUPLE STANDARDS:

An alternate approach to calibrating thermocouples and thermocouple materials, and one which gives a high degree of accuracy, is calibrating with a secondary standard that has similar FMF properties to those of the test element. That is, calibrating a KP element against a KP standard.

All thermocouple materials in the USA are referenced to a pure platinum element (PT-67) which is retained at the National Institute of Standards and Technology. It is important that regardless of the type of standard used that traceability to this NIST standard be accomplished.

CALIBRATION PROCEDURE:

The instruments mentioned previously as standards, fixed point cells, platinum resistance thermometers and liquid-in-glass thermometers can all be used to accurately calibrate thermocouples if the proper calibration procedures are followed. As previously mentioned, the two most common procedures are ASTM E207 & P220. We will now examine both more closely.

ASTM E207 WIRE TO WIRE CALIBRATION:

In order to achieve the maximum amount of accuracy when using base-metal standards, it is desirable whenever possible to make wire to wire readings, that is, reading the EMF developed between the thermocouple material to be calibrated and the base-metal standard of similar material.

Thermocouple material should be free of contaminants, insulated wire should be stripped of insulation. Insulation should be removed carefully in order to avoid cold working by nicking or stretching. Any one of these conditions could cause erroneous calibration results.

FLUIDIZED BEDS:

The fluidized bed is a unique method of providing closely controlled temperatures. The bath consists of a very fine mesh aluminum oxide, a heated chamber into which the medium is placed, and a means for slowly agitating the bath by introducing a flow of air. By careful control of heat input and air flow, temperatures of the bath can be controlled within close limits thereby producing isothermal conditions between the calibration standard and the test setup. Fluidized beds are useful for calibrating over the temperature range from ambient to 1600 degrees F.

STIRRED LIQUID BATHS:

Stirred liquid baths operate on the same principal as fluidized beds and are an excellent means of establishing closely controlled temperatures. Although stirred liquid baths using molten salts or liquid tin are available with a temperature range as high as 932 degrees F., the most common application is in the range of ambient to 500 degrees F. utilizing silicone oil as the bath material.

TUBE-TYPE HEATING ELEMENT FURNACE:

For temperatures above 500 degrees F. an electrically heated tube furnace is recommended. Tube furnaces operate in the range of 500 to 3100 Degrees. F.


Figure A:


Figure B:


Figure C:


Figure D:

 

 


Contact Us:  Seattle  ::  Spokane  ::  Portland  ::  Sacramento  ::  Boise
Copyright © 2002
All Rights Reserved
Last Updated: 19-Nov-2002