Home
Line Card
Product Category
Literature
Trade Show Schedule
Service & Calibration
Company Information
 


  
http://www.branom.com
sales@branom.com
800-767-6051
 

Home > Applications

Operations and Maintenance Manual
Gauges and Gauges mounted on Diaphragm Seals
(consult ANSI B40.l or Branom for additional data)

 

VACUUM, COMPOUND AND PRESSURE GAUGES

WARNING

Read before installing or using gauge. Misuse of this product may cause explosion and personal injury. Safety glasses are recommended when working with pressurized equipment.

INSTALLATION:

Never hand-tighten a gauge by its body; always use the wrench flats. Mount the gauge at or below eye level to reduce parallax errors (visual angle errors). Be sure the case is not distorted by mislocated or overtorqued mounting. To avoid contaminating the pressure measuring system in the gauge, leave a protective cap on the connection fitting until just before the line connection is made.

If a thread sealant, such as teflon tape, is used, be sure the sealant is compatible with the contact fluids. For example, never use teflon tape in the presence of silicon; use rector seal instead. Using the wrench flats, make the fitting up to proper tightness to prevent leakage. Do NOT, however, overtighten the fittings since they may then be damaged beyond repair. Do NOT use the gauge case to screw the gauge into place on the line fitting.

OPERATION: The most accurate reading is made by closing one eye, come up to indication point slightly from below to eye level, and then observing the position of the needle with the increments of the dial. For greater accuracy, mirror-dialed gauges help reduce the chances of parallax error. A zero-corrector calibration adjustment is a desirable feature for dry gauges (not available with liquid filled types). Mechanical or fluid vibrations, pressure spikes, may cause the needle to read off-zero. Calibration with portable pneumatic or hydraulic pump and test gauge is an inexpensive method of maintaining accuracy with a known accuracy and adjustable pressure settings. At least 10 points on the gauge should be tested while increasing from zero to full pressure and then while decreasing the pressure from the full scale reading to zero. A leak may be indicated by a continuously decreasing gauge reading, particularly if the pressure has been "trapped" by a valve. If this is the case, check all the components in the pressure system, including the fittings and the valve, for leakage. Some gauges appear to be out of calibration due to a static electrical charge. Removing the static charge with "Armor All" or other suitable cleaner should restore the gauge to proper calibration.

Pressure gauges must be correctly selected and installed to avoid possible injury or damage caused by misuse. Please note:

OXYGEN SERVICE: Gauges must be specially cleaned for oxygen service by a qualified laboratory, bagged and labeled for oxygen service. Branom offers this service. Repairs and testing of oxygen gauges must be done with clean air, nitrogen, or distilled water only. Never use any gauge on oxygen service that has been used on any other service, or tested with oil.

PRESSURE RANGE: Gauge dial range should be twice the intended operating pressure range. Never allow dynamic pressure to be applied beyond 90% of the scale. Please take into consideration possible spikes of overpressure; even those of short duration will cause impairment or failure. Branom does offer PRESSURE LIMITING VALVES, which can be set to guard against damage resulting from excess pressure. The PLV shuts off the instrument line automatically, and seals out pressure rises above the adjustable, pre-set value, and automatically restores the instrument line when the pressure falls below the setpoint. Gauges that shall see even occasional fluctuations between positive and negative pressure (vacuum) must be compound gauges, i.e. 30" Hg vacuum x 15 PSI or higher.

VIBRATION: Isolate gauge from sources of vibration to avoid wear to movement, which results in decreased accuracy. Mounting the gauge remotely from the vibration via 1/4" copper or SS tubing is best; when this is not possible, liquid filled gauges and the Vilter Stedigage devices have been used with some success.

TEMPERATURE: Temperature of the media entering the gauge should never exceed 1500F/650C (there are some exceptions, such as the Noshok 400 series @ 2600F). As in vibration, the best method to solve high temperatures is to run a copper or other material coil of 6' or longer, so that the temperature has been reduced before entering the gauge. Some diaphragm seals with or without remote capillary can sometimes also be used for this purpose; due to the expense and possible additional errors in accuracy, this is done usually only when solidification through temperature variations can take place.

Corrosion: All wetted parts of the gauge must be resistant to the process fluid utilized. Again, a diaphragm seal (available in plastic, for example) or an in-line isolator with the proper elastomer might be used with corrosive fluids, when gauges made of exotic metals or plastics are expensive or rare.

Pulsation: Excessive pulsations may cause substantial premature wear or work-hardening of the metallic parts and subsequent failure. Gauge readability may also be reduced dramatically due to pointer flutter. Internal spud restrictor screws, Liquid Filled Gauges, Porous disc and piston type pressure snubbers, and needle valves are recommended for this service. The needle valve is a particularly excellent pulsation dampener, as it can be throttled down as needed and also used as a shut-off valve. With porous disc snubbers, the gauge should be filled with a compatible glycerin or glycol or silicon, so that the disc can stay clean and porous as long as possible.

Liquid Filling for Gauges and Seals: Never use standard glycol or glycerin fills for applications involving strong oxidizing agents, including (but not limited to) oxygen, chlorine, nitric acid, and hydrogen peroxide. These substances can combine explosively with a standard fill. Typically, an inert liquid such as Fluorolube or Halocarbon is used with strong oxidizers. These, however, are never to be used with aluminum or magnesium.

AMMONIA Service: Wetted parts of carbon steel or stainless grades of wetted parts can be used with ammonia compounds. Gauges are available with corresponding temperature scales as well as pressure. Never use viton elastomers with ammonia.

Freon and Pressurized Nitrogen Service: Both of these gases are inert, and therefore brass, steel, monel or stainless grades of wetted parts can be used. High and Low Side Gauges are available with corresponding temperature scales as well as pressure for R12, R22, R502 and Rl34a.

STEAM SERVICE: Due to the high temperatures involved with steam, a steam syphon or pigtail filled with water should be attached to the gauge. A copper tube spiral could serve the same purpose.

DIFFERENTIAL: Indicating the difference between two pressures, the differential pressure gauge (DP Gauge) must be chosen not only for its materials, but also for the static pressure it will see. It is possible to have a few inches of water column differential on a 500 psi line. Some gauges can handle this high pressure, low differential service, some cannot. Some gauges, such as a Magnehelic, are designed only for air and gasses. Careful attention to detail will insure proper selection.

WATER LEVEL: Feet of Water Gauges are designed to be screwed into the side of a tank near the bottom and to measure the hydrostatic pressure in feet and inches. A Water Level Gauge, on the other hand, is a gauge designed to be installed in a well with a discharge column (tubing or pipe) and an airline of the same length. This gauge has an adjustable outer ring on the dial, calibrated counterclockwise, with two sets of figures. The inner readings are from 0-190 feet; the outer include 200-390 feet. The adjustable dial ring is set at the total number of feet of airline. Air pressure is then applied to the air line, purging the air line of water and causing the pointer to move away from the stop pin in a clockwise position. When the air line has been completely purged of water, the gauge hand will remain stationary, indicating the water level in feet from the Surface on the inner scale of the adjustable dial ring.

 

DIAPHRAGM SEALS & ISOLATORS

Filled system gauges and isolators, with or without trees (post for multiple instrument mountings), must be calibrated in the orientation necessary for the installation. It is recommended that testing under pressure be done 24-48 hours after filling and assembly, to insure that there are no leaks (316SS fittings, in particular, tend to leak over time, unless correctly torqued). Even minuscule leaks are catastrophic in filled systems.

Extreme care must be taken during installation to maintain the integrity of the seal between the gauge and the isolation device. Any movement of the connections, however slight, can cause a leak, and the device to "lose its fill". Excessive vibration, stepping or even leaning on the device, or personnel trying to turn the dial of the gauge for a better viewing angle, all cause catastrophic failures by causing leaks in the system.

Snubbers and needle valves should never be placed between the gauge and the diaphragm seal. Shutoff valves, either needle or ball valves (preferred), should be screwed into the process outlet of the diaphragm seal, and not be a part of the filled system. A snubber serves no purpose, and indeed, brings needless problems and expense to the gauge/diaphragm seal assembly. The use of a "gauge keeper" assembly, such as the Conoflow 100 A series #580280SF is recommended to reduce the possibility of gauge rotation and subsequent loss of fill.

The problems of standard in-line isolator ring assemblies can be eliminated by the use of Onyx brand patented system. The Onyx Isolator is filled at the factory with high viscosity silicon fluid. After filling, the Isolator Ring is permanently sealed a rubber "Module Seal", which enables the user to remove or replace or move the dial of gauges and other pressure instruments while the Isolator Ring is in the process pipe, without having to vacuum fill the ring or pressure instrument. No loss of fill, ever.

Pressure isolators frequently handle chemically reactive and abrasive fluids; it is reasonable to expect the elastomer sleeve to eventually wear out. Instrument oil must be compatible with the process liquid. Mixing incompatible fluids may cause a violent chemical reaction resulting in equipment damage or personal injury.

Make sure the isolator elastomer is compatible with the temperature, pressure and material being measured, and that the mounted instruments are corresponding in nature.


The Onyx Isolator Ring Out-Performs Conventional Isolator Rings And Standard Diaphragm Seals

Obtaining accurate pressure readings on pipelines is difficult because slurries, abrasives and solids clog gauges, switches and transmitters. Standard diaphragm seals - which are drilled or welded at a point on the pipeline - provide a temporary solution, but they are vulnerable to plugging by debris or deposits and stop working when slurry solidifies in the seal chamber. Conventional isolator rings consist of a rubber "inner tube" captured in a steel ring. The assembly is installed between flanges in the process pipe, with clear instrument fluid behind the rubber membrane transmitting pressure to the gauge. The inside diameter of the ring assembly matches the adjacent pipe, so the ring is continually cleaned by the motion of the process fluid. A common problem with this type of isolator ring is that any air in the instrument will compress when pressurized. This stretches the rubber membrane, causing accuracy to degrade or the instrument to malfunction. (An air bubble as small as a pea can make a sensor non-functional.) To counter this obstacle, isolator ring manufacturers use a vacuum pump to evacuate the air from the isolator, before injecting the instrument fluid. To remove the gauge for replacement or calibration, the entire ring and gauge assembly must be removed from the process pipe and the gauge and ring assembly again have to be drained, evacuated and refilled. The customer for this type of isolator ring will experience much downtime and expense. Some manufacturers try to circumvent this problem by adding a valve to the ring assembly. This holds the fluid in the ring, but attempting to fill the gauge by turning it upside down and pouring fluid into it traps an air pocket at the tip of the Bourdon tube. Also, attempting to remount the gauge by covering the connection with your finger while turning it upright and screwing it into the isolator is rarely successful.

The Onyx Isolator Ring Offers The Solid Solution Towards Solving Your Pressure Measurement Problems

The Onyx Isolator Ring provides a practical, cost-effective method for obtaining accurate pressure measurements on slurries, abrasives and hard-to-handle materials. The Onyx Isolator Ring is a new patented design which solves the problems associated with many of the isolator rings on the market today. The Onyx Isolator Ring is vacuum-filled at the factory with a high-viscosity silicone fluid; then, permanently-sealed with Onyx's Revolutionary "Module-Seal". With the Onyx Isolator Ring, there is no fill port anywhere on the assembly: gauges, switches and transmitters are supplied separately, are pre-filled, and have Onyx's special fitting attached. You have the option of selecting any combination of isolator ring and gauge and simply snapping them together. With the Onyx Isolator Ring, you are guaranteed hassle-free operation and added protection for your sensitive indication instrumentation.

 


Contact Us:  Seattle  ::  Spokane  ::  Portland  ::  Sacramento  ::  Boise
Copyright © 2002
All Rights Reserved
Last Updated: 19-Nov-2002